ALCOA+ Data Integrity in the Analytical Laboratory

What data, raw data and metadata mean, the nine ALCOA+ attributes written as paper and electronic controls that can be tested, where risk-based effort goes first, the data life cycle from acquisition to archive, hybrid systems, access control, and a self-inspection checklist for a small laboratory.

ALCOA+ requires that every record a laboratory relies on be attributable, legible, contemporaneous, original and accurate, and in addition complete, consistent, enduring and available, from the moment a value is generated until the record is destroyed under a retention schedule. The MHRA guidance defines data integrity as the degree to which data are complete, consistent, accurate, trustworthy and reliable, and the maintenance of those characteristics throughout the data life cycle 1. In practice each attribute is a control that can be tested on the laboratory floor: a named user behind every entry, an electronic record still readable in ten years, a chromatogram whose processing history can be reconstructed. This page converts the nine attributes into those controls.

The companion references on the laboratory notebook and on batch traceability describe two specific record types. This page is the framework above both: what counts as data, raw data and metadata; the attributes as paper and electronic controls; where risk-based effort goes first; the life cycle of a record; hybrid paper and electronic systems; the access and reporting culture the current guidance treats as part of integrity; which document governs which laboratory; and a self-inspection checklist. Worked examples refer to a UV assay of an invented decapeptide, AGSKLEYFVR, and to a generic chromatography data system; neither describes a real product or a real system.

Flat technical infographic of nine labelled-free hexagons arranged in a ring around a central circular arrow cycle divided into segments, representing the ALCOA+ attributes surrounding a data life cycle from acquisition to archive
The nine attributes surround the life cycle. Each attribute is a control that applies at every stage, from the first capture of a value to its archive and destruction.

Data, raw data, metadata

The guidance documents agree on three definitions, and most integrity failures trace back to a laboratory using one of the words loosely. Data are facts, figures and statistics collected together for reference or analysis, in any form. Raw data are the original records and documentation, retained in the format in which they were originally generated, paper or electronic, or as a true copy; where a basic instrument such as a balance or pH meter stores nothing and provides only a printed output, the printout constitutes the raw data 1. Metadata are data that describe the attributes of other data and give them context and meaning: structure, relationships, the identity of the user, the time stamp, the method version, and the audit trail itself 1.

TermDefinition in useIn the worked assay
DataAny recorded fact or figure, in any formatThe purity figure 98.4%; the sample mass; the peak table
Raw dataThe first capture, in its original format, or a verified true copyThe acquired data file from the detector; the balance printout; the handwritten dilution record
MetadataData describing the data: who, when, how, with whatUser ID, acquisition time, method and version, integration parameters, column serial, audit-trail entries
Static recordFixed content that cannot be interacted withThe balance printout; a PDF of the signed report
Dynamic recordContent the user can still interact withThe chromatogram, which can be re-integrated and re-reported
The three terms applied to one assay of the invented decapeptide.

The static and dynamic distinction is the one that decides most arguments about chromatography. The FDA guidance treats a chromatogram as a dynamic record because its content can still be reprocessed, and therefore a printout of it is not a complete copy of the raw data 4. A peak area reported without the integration events, the method version and the audit trail that produced it is a figure, not a record.

The nine attributes as testable controls

MHRA notes that the four additional attributes make explicit what was already implicit in the original five; PIC/S and WHO list the same nine 123. The table below gives, for each, the requirement, a paper control, an electronic control and the failure that most often appears in inspection findings.

AttributeRequirementPaper controlElectronic controlTypical failure
AttributableIt is clear who generated or changed the record and whenInitials and date on every entry; signature log linking initials to namesUnique user accounts; no shared logins; audit trail records user and timeGeneric 'analyst' account used by three people
LegibleReadable and permanent for the whole retention periodIndelible ink; single-line corrections; no overwriting or correction fluidFormats that remain readable after software upgrade; migration verifiedThermal printouts faded to blank in the archive
ContemporaneousRecorded at the time the activity is performedEntry made at the bench as each step is done; no transcription from scrapsSystem time stamps from a locked, synchronised clockWeights written up from memory at the end of the day
OriginalThe first capture, or a verified true copyThe bound notebook page or instrument printout, not a photocopyThe acquired data file retained; printouts are reports, not originalsChromatogram PDFs kept; data files deleted to save space
AccurateFree from error, reflecting what was observed, with corrections explainedCorrection keeps the original visible, with reason, initials and dateValidated calculations; audit trail captures change and reasonRe-integration until the result passes, with no reason recorded
CompleteAll data including repeats, reanalyses and failures retainedAborted runs and voided pages kept and cross-referencedNo deletion rights for analysts; all injections in the sequence retainedA 'test' injection of the sample before the official run
ConsistentSteps in the expected sequence, dates and times in orderPage numbering; sequential dates; date formats fixedSequence logs; time zone and clock fixed; no manual back-datingA review signature dated before the acquisition it reviews
EnduringPreserved for the retention period on durable mediaArchive-grade paper, bound; stored dry; controlled accessBackup verified by restore test; media refreshed before obsolescenceBackups never tested; drive fails and the restore does not work
AvailableRetrievable for review, audit or inspection within a defined timeIndex of notebooks; archive location recordedReadable copy of the software or exported open format kept with the dataData exist only on a decommissioned instrument's hard disk
ALCOA+ attributes as controls.

Data criticality and data risk

The guidance does not ask a laboratory to control every record to the same degree. The effort and resource applied to assure the integrity of data should be commensurate with the risk to product quality and balanced against other quality assurance demands 1. Two factors set that effort. Data criticality is the importance of the decision the data influence: a release result ranks above a training record. Data risk is the vulnerability of the data to deliberate or inadvertent alteration, deletion or recreation, and the probability that such a change would be detected 1. PIC/S frames the same assessment around the process, the degree of automation, the complexity of the system and the consistency of the output 2.

The data life cycle

MHRA defines the life cycle as all phases in the life of the data, from generation and recording through processing, including analysis, transformation or migration, use, retention, archive and retrieval, and destruction 1. Each stage has a characteristic failure and a characteristic control. The numbered map below follows one value, the sample mass of the invented decapeptide, from balance to archive.

  1. Generation. The balance prints 10.12 mg with date, time and balance identity. The printout is the raw data. Control: printer attached, print on every weighing, printout attached to the notebook page and cross-referenced.
  2. Recording. The analyst writes the mass, the balance ID and the sample label in the notebook at the bench. Control: contemporaneous entry; identical value to the printout; initials and date.
  3. Processing. The mass enters a locked spreadsheet that calculates the stock concentration, and the data system integrates the chromatogram against that concentration. Control: validated spreadsheet with protected formulae; integration method version-controlled; manual integration recorded with a reason.
  4. Review. A second person checks the raw data, the metadata and the audit trail against the reported result before approval, not the report alone. Control: review procedure names what is examined; the audit-trail review is documented.
  5. Reporting. The certificate or report states the result, the method version and the data file reference. Control: the report is traceable back to the data file and the notebook page.
  6. Use. The result drives a disposition decision. Control: the decision record cites the report, not a verbal figure.
  7. Retention. The notebook, printouts, data files and audit trail are kept for the retention period, accessible on request. Control: index; defined storage; backup restore tested at a set interval.
  8. Archive and destruction. At the end of active use the records move to controlled archive storage; destruction follows a documented schedule, is recorded and is approved. Control: archive log; destruction record; no destruction while any investigation or request is open.

Hybrid systems: printouts and electronic originals

A hybrid system holds the original record electronically and the review, approval or signature on paper. Most small laboratories run one, whether they call it that or not: a data system whose reports are printed, signed and filed. The guidance permits hybrids but treats them as a higher-risk configuration. The whole data set must be defined, the paper and electronic parts must be linked so that one cannot be reviewed without the other, and the electronic original must be retained for as long as the paper 12.

Access control, culture and the blame-free report

Three organisational controls recur across the guidance and are cheap to implement. Individual accounts: the FDA guidance states that shared login accounts are not acceptable, because actions cannot then be attributed to a single person 4. Separation of roles: the person who acquires and processes data must not hold administrator rights that allow the audit trail to be disabled, the clock to be changed or files to be deleted; Annex 11 requires physical and logical controls restricting access to authorised persons, with the creation, change and cancellation of access authorisations recorded 5. Audit-trail review: a documented review of the audit trail for each reportable result, by someone other than the person who generated the data, before approval 45.

Which guidance governs

Six documents cover the field. They agree on the attributes and differ in scope and depth. A laboratory that is not inspected under any of them still has no other agreed definition of trustworthy data to work to, and applying the controls proportionately costs little.

DocumentIssuer and yearScopeDistinctive content
Guidance on GxP data integrityMHRA, March 2018All GXP disciplines: GMP, GLP, GCP, GDP, GPvPDefinitions: data, raw data, metadata, true copy, life cycle, criticality and risk
PI 041-1 Good Practices for Data Management and IntegrityPIC/S, July 2021GMP and GDP inspectoratesLongest text; paper and electronic systems separately; organisational culture; inspector expectations
TRS 1033 Annex 4 Guideline on data integrityWHO, 2021Manufacturers and laboratories supplying WHO programmes; replaces the 2016 good data and record management textData governance, risk management, outsourcing, training
Data Integrity and Compliance With Drug CGMP: Questions and AnswersFDA, December 2018US CGMP for finished pharmaceuticals and APIsStatic versus dynamic records; shared logins; test injections; audit-trail review frequency
EudraLex Volume 4, Chapter 4 and Annex 11European Commission, January 2011EU GMP documentation and computerised systemsBinding requirements: audit trails, access, backup, archive, periodic evaluation
GLP Data Integrity, Advisory Document No. 22OECD, September 2021GLP test facilities for non-clinical safety studiesElectronic data capture; study director and quality assurance roles
The governing documents and their scope.

For a UK analytical laboratory the MHRA text is the primary reference and PIC/S the detailed companion 12. A laboratory supporting non-clinical safety work reads the OECD advisory document alongside them 6. The FDA question-and-answer format is the quickest route to a ruling on a specific practice, and its positions on printouts, shared accounts and trial injections are consistent with the European texts 4.

Self-inspection checklist for a small laboratory

Run the list once a year, and after any change of data system, instrument or staff. Record the answer to each item, the evidence examined and the action taken. An honest 'no' with a dated action is a compliant answer; an unexamined 'yes' is not.

  1. Is there a written definition of raw data for every instrument, stating whether the printout or the electronic file is the original?
  2. Does every person who touches data have an individual account, and is there a signature log linking initials to names for paper records?
  3. Are analysts prevented from deleting data, changing the system clock, or disabling the audit trail, and is the administrator account separate from any analytical account?
  4. Is the audit trail switched on for every data system, and is its review for each reportable result documented by a second person?
  5. Are all injections, including aborted, failed and repeated runs, retained and visible in the sequence record?
  6. Are calculation spreadsheets protected, version-controlled and validated against a worked example, with inputs traceable to raw data?
  7. Are notebook entries made at the bench, in ink, with corrections that leave the original legible and carry a reason, initials and date?
  8. Has a backup been restored successfully within the last twelve months, and is the restore recorded?
  9. Has every data-integrity concern raised in the period been recorded, investigated and closed, and was the person who raised it treated as having acted correctly?

References

  1. Guidance on GxP data integrity ('GXP' Data Integrity Guidance and Definitions, revision 1)Medicines and Healthcare products Regulatory Agency, 2018
  2. PI 041-1 Good Practices for Data Management and Integrity in Regulated GMP/GDP EnvironmentsPharmaceutical Inspection Co-operation Scheme, 2021
  3. TRS 1033 - Annex 4: WHO Guideline on data integrityWorld Health Organization, Technical Report Series, 2021
  4. Data Integrity and Compliance With Drug CGMP: Questions and Answers — Guidance for IndustryUS Food and Drug Administration, 2018
  5. EudraLex Volume 4 — Good Manufacturing Practice guidelines: Chapter 4 Documentation and Annex 11 Computerised SystemsEuropean Commission, 2011
  6. GLP Data Integrity — Advisory Document of the Working Party on Good Laboratory Practice (OECD Series on Principles of GLP and Compliance Monitoring, No. 22)Organisation for Economic Co-operation and Development, 2021