Skip to content
Peptides Info

Articles

80 reference articles, each cited to primary literature.

Analytical method design

Analytical calculations

  • Peptide Net Charge and pI: A Henderson-Hasselbalch Calculation

    The arithmetic that precedes buffer choice: the ionisable groups of a sequence, a stated pKa set, the fractional-charge formulas for acids and bases, a worked summation for an invented dodecapeptide at pH 3, 5, 7 and 9, bisection to the isoelectric point, why calculators disagree, and what the result means for solubility.

  • Relative Response Factors: Why Area % Is Not Weight %

    The correction behind an HPLC purity figure: response factor and relative response factor defined, RRF determination by impurity standard or slope ratio, the 1 ÷ RRF correction applied to a worked four-impurity chromatogram, the 0.8 to 1.2 window in which correction is commonly waived, and the ICH Q3A thresholds the corrected values are judged against.

  • Assay, Purity and Content: Three Numbers That Are Routinely Confused

    Purity is a ratio among related substances. Content is a mass fraction of the whole material on a stated basis. Assay is a measurement against a reference standard that produces a content. Definitions, the mass balance that links them, a worked lot and how each reads on a certificate.

  • Calibration Curves: Linearity, LOD and LOQ

    Building a calibration curve, judging linearity from residuals rather than r², and calculating detection and quantitation limits by the 3.3σ/S and 10σ/S conventions. A worked low-level curve and the rules for reporting results below the limit.

  • International Units: Why an IU Is Not a Mass and Cannot Always Be Converted

    An International Unit is a quantity of biological activity defined by a WHO reference preparation, not a quantity of matter. Where a mass conversion exists, why it exists; where it does not, why no factor can be written. Insulin, vitamins D and A, and heparin as worked cases.

  • Molar Mass, Moles and Converting a Mass to an Amount of Substance

    The conversion is one division. Getting it right depends on which mass goes into the numerator and which molar mass goes into the denominator. Definitions, sequence arithmetic, salt and water corrections, and a worked example on an invented decapeptide.

  • Significant Figures and Rounding: How a Result Should Be Reported

    The number of digits in a result is a claim about the method that produced it. Counting rules, rounding once and at the end, rounding against an uncertainty, comparing with a specification, and why 99.47%, 99.5% and 99% say different things.

Instrumentation

Method and records

  • Control Charts in the Lab: Trending SST and QC Results

    System suitability says whether one run passed. A control chart says whether the system is changing. Baseline construction, three-sigma limits, the Westgard rule set and where it fits chromatographic data, and a worked retention-time series.

  • Decision Rules and Guard Bands: Conformity With Uncertainty

    The step after the uncertainty budget and before the report: choosing, agreeing and recording the rule that turns a result with an expanded uncertainty into a statement of conformity under ISO/IEC 17025 clause 7.8.6. Simple acceptance, guard bands of w = kU, binary and non-binary statements and the probability of a false accept, worked for a generic purity specification.

  • Method Validation Basics for a Small Laboratory

    Specificity, linearity, accuracy, precision, range and robustness defined in plain terms — which of them an identity method needs, which a quantitative method needs, what full validation costs against verification, and a proportionate programme for a laboratory of three.

  • Out-of-Specification Results and What to Do About Them

    The difference between out of specification and out of trend, why retesting is never the first move, how to interrogate the method before the material, the averaging trap, and what to do when a supplier certificate contradicts your own result.

  • The Laboratory Notebook and Contemporaneous Records

    Why recording at the time is not the same as recording accurately later, the attributable-legible-original-accurate vocabulary, how to correct an entry without destroying it, bound versus electronic notebooks, and how long to keep them.

  • Uncertainty of Measurement in Practice

    Why every result is a range rather than a number, how Type A and Type B contributions differ, a worked uncertainty budget for a weighing and a dilution, the coverage factor and reporting convention, and what a purity figure quoted without uncertainty is actually worth.

  • Writing a Standard Operating Procedure: A Template That Survives Being Followed

    Scope, materials, procedure, acceptance, records — the five sections that carry an SOP, how to write steps somebody else can execute, how to version and review them, and a worked goods-in receiving skeleton to copy.

Quantification

  • qNMR for Peptide Content: Internal Standards and Limits

    Quantitative proton NMR assigns the absolute content of a weighed peptide by comparison with a certified internal standard, without a peptide reference material and without destroying the sample. The equation, a worked example, the acquisition settings that decide accuracy, and how the result compares with amino acid analysis and HPLC assay in an inter-laboratory study.

  • A205 and A214: Quantifying Peptides With No Trp or Tyr

    When the sequence has no tryptophan or tyrosine, A280 returns nothing. The far-UV alternative reads the peptide bond itself. Coefficients at 205 and 214 nm, a worked calculation for a generic decapeptide, and the buffer, counter-ion and stray-light constraints that make this harder than A280.

  • Amino Acid Analysis as a Reference Method

    The only routine assay that answers how much peptide is in a weighed solid. Hydrolysis and what it destroys, derivatisation and separation, what it costs, and what its absence from a certificate leaves undetermined.

  • Counter-Ion Content: Trifluoroacetate and Acetate

    Synthesised peptides arrive as salts, and the salt is not a trace. How much mass the counter-ion carries, how it is measured, when exchange is worth requesting, and where it sits in the accounting that reconciles gross weight to peptide.

  • Gravimetric and Volumetric Preparation Error

    Balance resolution against pipette tolerance, the volume below which weighing wins, the density and temperature terms that convert between them, and how error behaves as it passes down a dilution series.

  • Karl Fischer Titration and Residual Moisture

    Water is the largest non-peptide term nobody measures. How the titration works, what volumetric and coulometric modes are each good for, where the sub-three-per-cent convention came from, and what a moisture figure on a certificate does and does not describe.

  • UV Absorbance at 280 nm: Concentration Without a Balance

    An optical measurement of peptide concentration that is blind to counterion, water and salt — and that returns nothing at all for a sequence carrying no aromatic residues. Coefficients, path length, interference and the failure cases.

Degradation

Stability study design

Contamination control

Preparation

Storage

Transport standards

Verification

Data integrity

  • Audit Trail Review for Chromatography Data Systems

    What a chromatography data system audit trail records, how to enable and lock it, a checklist of what to look for in a review, who should review, how often, and what to write down. A procedure with sign-off fields.

  • ALCOA+ Data Integrity in the Analytical Laboratory

    What data, raw data and metadata mean, the nine ALCOA+ attributes written as paper and electronic controls that can be tested, where risk-based effort goes first, the data life cycle from acquisition to archive, hybrid systems, access control, and a self-inspection checklist for a small laboratory.

  • Validating Laboratory Calculation Spreadsheets

    A validation protocol for a spreadsheet that calculates purity, content, dilutions or uncertainty: requirements, design controls, a test-case table of nominal, boundary and invalid inputs, version and change control, and the status of printed outputs.

Formulation and lyophilisation

  • Lab Freeze-Drying: Tg', Collapse Temperature and the Cycle

    How Tg' and the collapse temperature are measured, how they set the product-temperature target, and how shelf temperature and chamber pressure follow from it. A cycle-design procedure for a laboratory freeze-dryer with a parameter table and a run record.

  • Lyophilisation Excipients: Bulking Agents and Lyoprotectants

    What mannitol, glycine, sucrose, trehalose and buffer salts each do in a freeze-dried peptide formulation, how each behaves on freezing and drying, what goes wrong with each, and a procedure for reading excipient content from a label and a certificate of analysis.

  • Cake Inspection: Collapse, Meltback and Reconstitution Time

    A bench procedure for inspecting a lyophilised cake: lighting and background, a defect vocabulary, a 1 to 5 grading scheme, what each defect says about moisture and drying temperature, the timed dissolution test, and the record to keep.